iFlash 1200 Maintenance Examination

Total Score: 100 Points | Total Questions: 21 | Time: 90 MinutesPlease fill in your name, region and test date before starting the exam. Answer each question according to the maintenance specifications of iFlash 1200, wish you good performance!

Q1:Name

填空1

Q2:Company

填空1

Q3:Region (Country)

填空1

Q4:Date

日期

Q5:For a unit shipped to Country A in January 2026, received in April, and installed in October, what is the estimated warranty expiration date?

2026.12
2027.7
2027.11
2028.4

Q6:Among the following customer complaints, which replacement parts are able to apply in-warranty replacement (the machine is still in-warranty)?

Functional failure of the peristaltic pump motor
Rupture of the peristaltic pump tubing due to aging
Clogging of the filter in the tube from outer tank to inner tank. Caused by water quality issues
Bending of the magnetic separation aspiration probe during operation

Q7:An instrument was shipped in March 2025, installed in November 2025, and the installation report was feedback in the same month. When does the warranty period for this instrument expire?

2026.11
2026.9
2027.1
2027.9

Q8:When a end-user is running Vitamin D samples and infectious disease samples(Like HIV, HbSAg...). During the training after installation , you should require them to perform aspiration probe maintenance at least at a frequency of

Every 3 months
Every 1 months
Every week
Every 3 days

Q9:When cleaning the sample probe or reagent probe, ( ) can be used.

Pure water-moistened wipes
75% alcohol-soaked tissue or gauze
Dry paper towels
Regular cleaning cloth

Q10:Required cleaning solution for Trigger Tip

75% alcohol
Purified water
Wash Buffer
Dry paper towel

Q11:Minimum inner tank cleaning frequency

1 months
3 months
6 months
12 months

Q12:Minimum outer tank cleaning frequency

1 months
3 months
6 months
12 months

Q13:When disassembling the iFlash 1200 magnetic separation aspiration probe, which screws should be removed?[图片]

A
B
C
D

Q14:To secure an extended warranty period, you should

填空1

Q15:The standard replacement interval for peristaltic pump tubes is

填空1

Q16:The standard replacement interval for filters is

填空1

Q17:Components requiring EVERY 6 months replacement have:

填空1

Q18:Components requiring cleaning & lubrication:

填空1

Q19:The peristaltic pump tube is broken of an instrument installed just a year ago, causing water ingress and short circuits that damaged other components. However, no maintenance records for this unit can be found in either the end-user site or distributor’s management system. The parts damaged due to peristaltic pump tubing leakage are eligible for in-warranty replacement.

True
False

Q20:Engineers must perform regular instrument maintenance to ensure performance, with a frequency not lower than the minimum required standard.

True
False

Q21:At an end-user site, poor water quality caused frequent clogging of the instrument’s filters. The user independently installed a non-specified filter model. Although the instrument initially functioned normally, the filter detached due to clogging one month later, causing a fluid leak that damaged several motors. The instrument is confirmed to be within the warranty period. The damaged motors are able to apply for warranty replacement under this claim.

True
False

Q22:An operator noticed dust and dirt around the RV port on the instruments shell and cleaned it with using a clean tissue and 75% alcohol. This cleaning procedure is correct.

True
False

Q23:When replacing the peristaltic pump tube of iFlash 1200, only need to replace the tube

True
False

Q24:Sample probe cleaning precautions:(7)

填空1

Q25:Precautions during magnetic separation aspiration problems disassembly?(10’)

填空1

Q26:Sample probe cleaning precautions(10’)

填空1